Blood pressure monitor error codes E1–E5: what each code means by brand
Blood pressure monitor error codes E1 through E5 are brand- and model-specific rather than universal across devices. In the official OMRON BP7365CAN manual (checked 2026-11-27), E1 points to cuff placement, a loose air plug, or an air leak; E2 means movement or systolic pressure above 210 mmHg; E3 is over-inflation; E4 is movement vibration; and E5 is an undetected pulse, while A&D Medical UA-651 monitors instead display Err, Err CUF, E, Err E, or Err 9.
General information, not medical advice. No medical review is claimed.
How can I use the free worksheet?
Print this page or download the blank text sheet. Keep the sheet with your original record. Use it to explain one question to your care team.
Error code troubleshooting and support sheet (blank template)
- □ Date and time: ________________________
- □ Monitor brand and exact model number (from bottom/back label): ____________________________________
- □ Exact error code shown (E1–E5, Er, Err, Err CUF, E, Err E, Err 9): _______________________________
- □ Air plug seated firmly and tubing straight (not pinched): yes / fixed
- □ Cuff placed on bare upper arm above elbow bend, snug wrap: yes / fixed
- □ Electrical devices at least 12 inches (30 cm) away: yes / fixed
- □ Quiet 5-minute rest, arm at heart level, still and silent: yes / fixed
- □ Fresh batteries installed or 30-second battery reset completed: yes / not needed
- □ Outcome after one repeat (or support/clinician note): ____________________________________________
The sheet has no score. It does not collect or send health details.
| Display code (brand/model) | What the official manual says it means | First fix step from the manual | Source & check date |
|---|---|---|---|
| E1 (OMRON BP7365CAN) | START/STOP pressed with cuff off; air plug loose; cuff not applied correctly; or air leaking from the cuff | Turn off, push air plug firmly into monitor, rewrap cuff correctly on bare upper arm; replace cuff if leaking | OMRON BP7365CAN manual p. EN37 (checked 2026-11-27) |
| E2 (OMRON BP7365CAN) | Moving or talking prevented sufficient inflation, OR systolic pressure is above 210 mmHg | Sit still and do not talk; if E2 repeats and you suspect very high BP or have symptoms, contact a clinician | OMRON BP7365CAN manual p. EN37 (checked 2026-11-27) |
| E3 (OMRON BP7365CAN) | Arm cuff inflated above the maximum allowable pressure | Do not touch the cuff or bend/pinch the air tube while measuring; check manual inflation instructions if used | OMRON BP7365CAN manual p. EN37 (checked 2026-11-27) |
| E4 (OMRON BP7365CAN) | Moving or talking during measurement created vibrations that disrupted the reading | Rest arm at heart level, remain completely still, and do not talk during inflation or deflation | OMRON BP7365CAN manual p. EN38 (checked 2026-11-27) |
| E5 (OMRON BP7365CAN) | Pulse rate was not detected correctly | Reapply the arm cuff properly on bare skin above the elbow bend, sit upright, and remain still | OMRON BP7365CAN manual p. EN38 (checked 2026-11-27) |
| Er (OMRON BP7365CAN) | The monitor itself has malfunctioned | Press START/STOP once more; if Er remains, try a 30-second battery reset (p. EN41) or contact OMRON support | OMRON BP7365CAN manual pp. EN38, EN41 (checked 2026-11-27) |
| Err (OMRON BP7365CAN) | Monitor cannot connect to a smart device or transmit data correctly | Follow app instructions; your on-screen reading is separate from Bluetooth sync. Contact support if Err persists | OMRON BP7365CAN manual p. EN38 (checked 2026-11-27) |
| Err (A&D Medical UA-651) | Movement made BP unstable, systolic and diastolic are within 10 mmHg, or pressure did not rise during inflation | Sit still for movement errors; reapply cuff snugly on bare upper arm if pressure failed to rise | A&D Medical UA-651 manual p. E-6 (checked 2026-11-27) |
| Err CUF (A&D Medical UA-651) | The arm cuff is not applied correctly | Remove and reapply the cuff snugly on bare skin above the elbow bend, check hose connection, and retry | A&D Medical UA-651 manual p. E-6 (checked 2026-11-27) |
| E (A&D Medical UA-651) | Pulse display error — pulse was not detected correctly | Reapply the cuff correctly over the upper arm and remain still during a calm repeat attempt | A&D Medical UA-651 manual p. E-6 (checked 2026-11-27) |
| Err E or Err 9 (A&D Medical UA-651) | Device internal error | Remove batteries, press the power button once, reinstall batteries; if the error persists, contact the dealer | A&D Medical UA-651 manual p. E-6 (checked 2026-11-27) |
How can I use this record checklist?
- Note the exact code on screen and leave the blood pressure fields blank in your log — an error code is not a reading.
- Check the air plug and rubber tubing: push the plug firmly into the monitor port and straighten any kinked or pinched tube.
- Remove the cuff and rewrap it on bare skin directly above the bend of the elbow so it sits snugly without pinching or slipping.
- Move phones or other electrical devices at least 12 inches (30 cm) away from the monitor.
- Rest quietly for 5 minutes with back supported, feet flat, legs uncrossed, and your arm supported on a table at heart level.
- Take one calm repeat reading without moving or talking. If a hardware code (Er, Err E, or Err 9) repeats after a fresh-battery reset, stop retrying and contact customer support.
Home measurement reference: American Heart Association. Use your clinician’s individual instructions.
Why does the same error code mean different things on different brands?
Home blood pressure monitors do not share one industry-wide error-code numbering standard. On an OMRON BP7365CAN upper-arm monitor, E1 through E5 cover cuff/air-plug problems (E1), movement or systolic pressure above 210 mmHg (E2), over-inflation (E3), movement vibration (E4), and undetected pulse (E5). Other brands assign those same numbers to different faults — or skip E1–E5 altogether, as A&D Medical does on the UA-651 with Err, Err CUF, E, Err E, and Err 9.
Before troubleshooting, check the model sticker on the bottom or back of your monitor so you match the exact code to your device manual. If you keep a paper or app diary, follow our blood pressure monitor error log guide so a failed inflation never gets mixed into your weekly average.
How do you tell a quick cuff or posture error from a hardware fault?
Most home errors happen because the cuff shifted, the air plug worked loose, or your arm moved during inflation. On OMRON BP7365CAN, E1, E3, E4, and E5 all point to setup, tubing, cuff fit, or movement rather than a broken pump. On A&D UA-651, Err, Err CUF, and E point to movement, cuff placement, or pulse detection.
Hardware and connection errors use separate labels. OMRON reserves Er for an internal monitor malfunction and Err for a smart-device Bluetooth transfer failure. A&D reserves Err E and Err 9 for an internal device error. If a hardware code returns after installing a full set of fresh batteries and resetting the unit, stop taking repeated inflations and contact the manufacturer or dealer.
When should you stop retrying and call a clinician instead of troubleshooting?
A repeated E2 on an OMRON BP7365CAN is a critical exception to routine cuff troubleshooting: page EN37 explicitly notes that E2 can occur when systolic pressure is above 210 mmHg and the monitor cannot complete automatic inflation. Repeatedly re-inflating the cuff on a sore arm will not resolve genuinely very high blood pressure.
Per American Heart Association guidance, sit quietly for 5 minutes with proper posture and bare-arm cuff placement before a repeat attempt. If your monitor keeps failing to inflate and you suspect very high blood pressure — or if you have chest pain, shortness of breath, numbness, weakness, vision changes, or severe headache — seek medical care promptly rather than continuing to troubleshoot the device. This guide is for general information and is not medical advice.
What setup checks clear most E1–E5 and Err CUF codes on the first retry?
Start at the port: push the air hose connector straight into the monitor until it seats firmly and inspect the tubing for cracks or folds. Next, check cuff sizing and placement on bare skin directly above the elbow bend — a cuff that is too loose leaks air and slides, while a cuff that is too small can pinch or trigger over-inflation. See our blood pressure cuff fit guide and cuff size chart if your index marker falls outside the printed OK zone.
Check power and interference too. Weak batteries can stall inflation mid-cycle (see our low battery readings guide), and OMRON advises keeping phones or other electrical devices at least 12 inches (30 cm) away during a reading. Once you finish fixing setup, rest calmly for 5 minutes with your arm supported at heart level before pressing Start.
What else do people ask?
What does E1 mean on an OMRON blood pressure monitor?
In the OMRON BP7365CAN manual (p. EN37), E1 means the Start/Stop button was pressed with the cuff not applied, the air plug is not fully plugged in, the arm cuff is not applied correctly, or air is leaking from the cuff. Turn the unit off, seat the air plug firmly, and rewrap the cuff snugly on bare skin; replace the cuff if it leaks.
What does E2 mean on an OMRON blood pressure monitor?
In the OMRON BP7365CAN manual (p. EN37), E2 means moving or talking prevented the cuff from inflating enough, OR your systolic blood pressure is above 210 mmHg so automatic inflation could not complete. Remain completely still for one calm repeat; if E2 persists or you have symptoms, contact a healthcare professional.
What is the difference between Er and Err on an OMRON monitor?
On the OMRON BP7365CAN (p. EN38), Er means the monitor itself has malfunctioned, while Err means the monitor could not connect to a smart device or transmit data to the app. A Bluetooth Err message does not mean your on-screen blood pressure reading failed.
What does Err CUF mean on an A&D Medical blood pressure monitor?
In the A&D Medical UA-651 manual (p. E-6), Err CUF means the arm cuff is not applied correctly. Remove the cuff, wrap it snugly on your bare upper arm above the elbow bend, confirm the hose is seated cleanly, and repeat the reading while sitting still.
How do you reset a blood pressure monitor after an internal error code?
For OMRON BP7365CAN general errors (p. EN41), turn the unit off and back on; if the error continues, remove all batteries, wait 30 seconds, and reinstall them. For A&D UA-651 Err E or Err 9 (p. E-6), remove the batteries, press the power button once, then reinstall batteries. Contact customer service or the dealer if the internal error returns.
Should I write an error code into my blood pressure log?
Record the date, time, and exact error code in the notes column, and leave the systolic and diastolic boxes blank. Never guess or enter an estimated number when the monitor displays an error.
Can the Insider Health habits quiz diagnose why my monitor shows an error?